GE Datex-Ohmeda Aisys Carestation Reference Manual page 249

Anesthesia machine
Table of Contents

Advertisement

Error Log Entry
Action/Troubleshooting
O2 PROP VALVE FAIL
Reboot System. If problem continues, Replace the Mixer.
O2 SENSOR FAILURE
Calibrate O
If calibration fails, replace O
If calibration continues to fail, wait 90 minutes and repeat calibration.
If calibration fails after 90 minute, replace VIB.
ON/STANDBY SWITCH
TO STANDBY
No Service Action Required.
OUTFLOW SCAV 10VA
POWER ERROR
If the P2 indicator never lights or lights only briefly,
• proceed to interconnect fault isolation procedure (Section 7.10.12).
If the P2 indicator remains lit for more than 2 seconds,
• isolate the failed valve (outflow or scavenging) circuit with manual valve controls.
Replace and retest in the following order (Also see Section 7.10.):
• Valve
• ADB
OUTFLOW ZERO POINT
ERROR
Replace and retest in the following order (Also see Section 7.10.):
• Outflow Zero Valve
• Flowmeter Block
• ADB
OUTPUT FLOW LIMIT
REACHED
Indicates commanded agent flow could not be achieved because the vaporizer reached its maximum
flow capability. This occurs at high flow and agent settings, primarily with Sevoflurane. The user sees an
alarm message advising them to reduce flows and the agent monitor may show an under delivery.
M1046983 04/08
Alarm Text
Condition (Basic info)
Mixer status bit
STS_CH1_PROPN VALVE FAIL
indicates proportional valve
failure.
Replace O
sensor
O
< 5%
2
2
Sensor.
2
Sensor.
2
Turn switch on to
On/Standby switch
continue use
transitions from On to
Standby.
Vaporizer Failure
Overcurrent condition
detected by the AC. Circuit
disabled.
Vaporizer Failure
Output flowmeter measured
value during zeroing is out of
range.
Cannot deliver agent
Commanded cassette flow >=
setting at set flow
6.0 L/min for > 10 seconds
OR
Commanded flow >= 4.0 L/
min and flow valve is at max
for > 10 seconds.
7 Troubleshooting
Priority Source Enabling Criteria
Medium
AC,
Mixer
Low
AC,
Galvanic O
Vent
connected
High
PC
System state is
Therapy and Power
Controller is
communicating with
DC
AC
AC
Low
AC
Agent delivery on
sensor
2
7-67