12.7 Power Diagnostics menu
12.7.1 Power Diagnostics
M1046983 04/08
The Power Diagnostics menu includes the following menu items:
• Power Schematic (Section 12.3)
• Power Board (Section 12.7.1)
• Anesthesia Control Board Power (Section 12.7.2)
• Mixer Board Power (Section 12.7.3)
• Vent Interface Board Power (Section 12.7.4)
• Display Unit Power (Section 12.7.5)
• Vaporizer Power (Section 12.7.6)
Power Board
Power Board
Label
12Vdc Supply
3.3Vdc Supply
1.5 Vdc Supply
Battery Connected
Calc Battery Time
Battery 1 Volts
Battery 2 Volts
Battery Current
Battery Status
<*> Date battery Tested
Last Full Discharge
Board Temperature
Fan Speed
Fan 1 Voltage
Fan 1 Status
Fan 2 Voltage
Fan 2 Status
<*> Date battery Tested = date of Last Full Discharge (refer to section 6.10).
Value Format
XX.XX
X.XXX
X.XXX
Yes, No
XX
XX.X
10.00 to16.50 (black)
<10.00 (red)
>16.50 (red)
X.XXX
Fail
Bulk Chg
Over Chg
Float Chg
Trickle Chg On
Discharge
--/---/----
XX
≤75C OK (black)
>75C Fail (red)
Slow, Fast
XX.XX
OK, Fail
XX.XX
OK, Fail
12 Service Application
Units
Range
Vdc
11.70 to 12.30
Vdc
3.201 to 3.399
Vdc
1.450 to 1.550
Min
00 to 30
Vdc
10.00 to 16.50
A
- 11.000 to 4.000
Min
C
max 64
F
max 147
Vdc
Fast 11.52 - 12.48
Slow10.08 - 10.92
Vdc
Fast 11.52 - 12.48
Slow10.08 - 10.92
12-13