To Replace The Lcd Display - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys
9.11.5 To replace
the LCD display
9-26
1. Disconnect the following cables:
• Right membrane switch flex-cable at ZIF (zero insertion force) connector (A)
• Encoder assembly cable (B)
• Lower membrane switch flex-cable at ZIF (zero insertion force) connector (C)
• Fan cable (D)
• LCD cable (E)
• Left membrane switch flex-cable at ZIF (zero insertion force) connector (F)
2. Remove the ten screws (circled) that hold the mounting plate to the front enclosure.
A
B
3. Remove the mounting plate assembly from the front enclosure.
4. Disconnect the backlight harnesses (G) from the inverter boards.
5. Slide the grommet (H) out of the mounting plate slot (transfer to new LCD).
C
D
04/08 M1046983
F
E