Evap Scavenger Path Testing - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys

7.14 eVap Scavenger Path Testing

7-102
1. Connect and establish communication with the PC Service Application.
File
Preferences
2. In the
>
Gas Delivery Schematic
3. On the
• Mixer O2 Flow to 0.50 l/min.
4. Insert a de-pressurized Test Cassette.
Vaporizer Schematic
5. On the
• Inflow valve to Off
• Outflow valve to Off
• Scavenging to On
• Prop Flow Valve DAC Value to 0
• Cassette Power to Off
• Zero Valves Auto box checked ( )
6. Record the C C C C a a a a s s s s s s s s e e e e t t t t t t t t e e e e P P P P r r r r e e e e s s s s s s s s u u u u r r r r e e e e value.
7. Set the Inflow valve to On.
8. After the reading stabilizes, re-record the C C C C a a a a s s s s s s s s e e e e t t t t t t t t e e e e P P P P r r r r e e e e s s s s s s s s u u u u r r r r e e e e .
9. Subtract the second reading from the first reading.
a. If the difference between the two readings is less than or equal to 1.00 kPa, there
is no restriction in the scavenger path. No action is required.
b. If the difference between the two readings is greater than 1.00 kPa, disconnect
the eVap scavenger tubing (blue tubing) and repeat the test.
c. With the scavenger tubing disconnected, if the pressure difference is less than
1.00 kPa, inspect the scavenger tubing for kinks and replace as necessary. If no
kinks or restrictions are noted, inspect the Scavenger Downtube (1407-3904-
000) for restrictions, replace as necessary.
d. If the pressure gradient remains greater than 1.00 kPa, troubleshoot or replace
the eVap.
menu set the Gas Supply Pressure Units to kPa.
, set the following:
, set the following:
04/08 M1046983