9.8 Servicing Aladin
Note
9.8.1 Emptying an
Aladin
cassette
2
9-16
cassettes
2
Before service or returning a cassette to the manufacturer, make sure that the cassette
is empty. All Aladin
cassettes must be emptied before shipping. Package and ship the
2
cassettes in a suitable container.
1. Connect an empty bottle to the filling system and hold it tight.
2. Turn the Aladin
cassette so that agent flows into the bottle and wait until the
2
cassette is empty.
3. To get the maximum amount of agent out, rock the cassette from left to right and tip
it forward and back several times.
4. Remove the bottle before returning the cassette to the horizontal position.