4.5.1 Spiro Calibration

Instructions
Spiro Calibration menu
M1046983 04/08
The Spiro Calibration instructions appear when the focus is on Spiro
Calibration menu item.
To display TV data:
1. Connect a spirometry sensor to MGas.
2. Push the ComWheel to continue.
3. Select the correct spirometry sensor type (Adult or Pedi).
4. Use the test device to deliver a known TV through the sensor.
5. If necessary, repeat steps 1-4 for the other type of sensor.
Menu Item
Values
Sensor Type
Insp Gain
TVinsp
Exp Gain
TVexp
4 Install/Service Menus
Comments
Adult or Pedi
Select Sensor type: Adult (Dlite) or
Pedi (Pedilite).
1000
_ _ _
ml
1000
_ _ _
ml
4-19