3.4 Bellows drop test

3.5 Backlight test

M1046983 04/08
Note: This check is NOT performed during the "System Check." This test is
used to aid in troubleshooting if the "System Check" fails.
1. End a case.
2. Set the Bag/Vent switch to Vent.
3. Occlude the patient Y piece.
O2 Flush
4. Push the
button until the bellows is full.
5. After the initial drop, if the bellows falls more than 100 ml/min,
it has a leak.
Main Menu
1. Push the
2. Select Calibration.
3. Select Backlight Test.
4. Select Start Test.
5. The display will show the test running on light 1 and then on light 2. If the
display goes completely blank or flickers during the test, one of the lights
has failed.
key.
3 Checkout Procedure
3-7