7.10.6 eVap Pressure troubleshooting

Contact Technical
Support for advice.
21
Flowchart
7-86
Start
Within the PC Service Application, with the
Inflow Valve and Scavenger Valve open,
establish 500 ml flow from the Mixer
(Section 12.3.4) (Section 12.3.2)
Subtract Mixer Outlet pressure from Cassette Pressure.
Yes
CASSETTE PRESSURE ERROR
CASSETTE PRESSURE ERROR
FLOW MANIFOLD EEPROM FAILURE
VAP FLOW METER BLK COMPAT FAIL
.
Is the difference between the
Mixer Outlet and Cassette
within -0.8 and 1.4 PSI?
No
Error Log Entry
Replace the
Flowmeter
Subassembly.
04/08 M1046983