Replace the relief valve
• Hold the backpressure plug (
pliers while turning the relief valve (
counterclockwise to remove it.
• If you are replacing the relief valve and the
backpressure valve (
), remove the relief valve
13
and the backpressure plug as a unit before
replacing the relief valve.
Replace the backpressure valve
• Loosen the mounting screw (
• Pull on the backpressure plug as you continue to
loosen the mounting screw until both the screw and
plug come loose.
• Use an angled probe to grab the backpressure
valve (
) and pull it out of the manifold.
13
(If the machine is on, hold a hand over the
backpressure valve bore and push the alternate O
button; the backpressure valve may pop out. Push
the alternate O
button again to disable flow.)
2
• Inspect and replace the plug o-ring if required.
• Insert the backpressure valve into the manifold; note the direction (flow arrow should point out of the manifold).
• Using only a finger, push the backpressure valve into the bore.
• Position the plug assembly and the mounting screw together in place.
• Tighten the mounting screw while pushing down on the plug assembly until it is fully seated.
To replace the Flowmeter subassembly
• Refer to the Installation Instructions included in the
Flowmeter Subassembly Kit.
Replace the inflow check valve assembly
• Refer to the Installation Instructions included in the
Inflow Check Valve Assembly Kit.
To replace the valve block subassembly
• Refer to the Installation Instructions included in the
Valve Block Subassembly Kit.
M1046983 04/08
) with non-marring
11
)
12
).
14
2
14
12
11
9 Repair Procedures
12
11
13
9-15