9.14.3 Replace
the inlet manifold
M1046983 04/08
1. Remove the rear panel (Section 9.3).
2. Disconnect the tubing from the manifold outlet(s).
3. Disconnect the transducer harness.
4. Remove the two screw that hold the manifold to the side extrution.
3
4
5. Transfer the following items to the replacement manifold or install new as required.
• pipeline check valve (1)
• inlet filter (2)
• inlet fitting (3) and o-ring (4)
• relief valve (5)
6. Transfer the pressure transducer to the new supply (Section 9.16).
• Ensure the o-ring is in place.
• Install the transducer.
7. To reassemble, perform the previous steps in reverse order.
8. Perform the checkout procedure (Section 3).
1
2
9 Repair Procedures
5
9-41