Aisys
9.19 Replace the APL valve
9-48
1. Remove the ABS breathing system.
2. The APL valve is held in place with a
spring and a retainer (A) that snaps into
a recess in the lower body of the APL
valve. To release the retainer, place an
appropriately sized straight blade
screwdriver into the housing cutout (B).
Twist the screwdriver to release the
retainer.
3. Place the new APL valve into position
with the setting indicator facing to the
front of the machine.
4. Place the spring into the retainer.
5. While holding the APL valve tight to the
housing, snap the spring and retainer
onto the valve body from below.
6. Reinstall the ABS breathing system.
7. Perform the checkout procedure
(Section 3).
B
04/08 M1046983
A
Front