Flush Flow (With Acgo) - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys
2.12.5 Fresh gas and O
To ABS (Circle)
breathing system
EV
1
ABS
ACGO
Selector
Switch
Figure 2-45 • Fresh gas and O
2-58

flush flow (with ACGO)

2
Fresh gas (1) flows from the electronic vaporizer (EV) to the ACGO Selector Switch.
With the ACGO Selector Switch in the ABS position, fresh gas flow is channeled to the
breathing system through port 3.
The output of the O
Flush regulator (2) is channeled to the O
2
When activate, O
flush flow joins the fresh gas flow in the ACGO Selector Switch.
2
To
Scavenging
From
Mixer
flush flow (to ABS)
2
From
O
Supply
2
To
Mixer
2
Absorber
2
5.4 psi
O2 Flush
Switch
ACGO
Selector
Switch
1
ACGO
Variant
Flush valve.
2
O2
Sensor
3
Insp Flow
1
Paw Trans
ACGO
04/08 M1046983

Advertisement

Table of Contents
loading

Table of Contents