Evap Adb Troubleshooting - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

7.10.9 eVap ADB troubleshooting

ADB Related
DCB RAM ERROR
PORT ZERO READ BACK FAIL
ADB VOLTAGE ERROR
VAP_PORT_ZERO_READ_BACK_FAIL
"Insert Cassette" Message
Replace the ADB Board
for any of the above errors.
24
Flowchart
M1046983 04/08
Is the Error entry
ADB
"Cassette Retention",
Cassette Retention
Cassette Retention Related
Cassette Not Latching
Replace:
• Cassette Handle
• Cassette Bay
Start
"ADB",
or "Agent Level"
related?
Agent Level
7 Troubleshooting
Symptom
Cassette Not Latching
No Liquid Level Indicator
Error Log Entry
ADB VOLTAGE ERROR
AGENT LEVEL LOST
AGENT LEVEL OVER RANGE
AGENT LEVEL UNDER RANGE
DCB RAM ERROR
INSERT CASSETTE
PORT ZERO READ BACK FAIL
VAP PORT ZERO READ BACK FAIL
Liquid Level Related
Does not display Liquid Level with any
Aladin II cassette or DES Aladin I.
Replace:
• Cassette Interface Board
• the ADB Board
7-89

Advertisement

Table of Contents
loading

Table of Contents