9.3 How to remove the rear panels

9.3.1 To remove
the rear upper
panels
9.3.2 To remove
the lower access
panels
9-6
To access components in the upper electronic enclosure from the rear of the machine,
you must remove the rear cosmetic panel and an inner enclosure cover.
To access components in the lower electronics enclosure, you must remove the lower
access panel.
1. Bleed all gas pressure from the machine (Section 9.2).
2. Ensure that all cylinder and pipeline pressures read zero before proceeding.
3. Disconnect all electrical cables.
4. To remove the rear cosmetic panel, fully loosen the five captive screws that hold the
panel in place. Remove the panel.
5. To remove the inner access panel, remove the 18 mounting screw around the
periphery of the panel. Disconnect the fan cable to remove the panel.
1. Disconnect the power cord from the AC mains supply.
2. Bleed all gas pressure from the machine (Section 9.2).
3. Ensure that all cylinder and pipeline pressures read zero before proceeding.
4. If present, remove the inboard cylinders.
5. To remove the lower access panel, remove the 11 mounting screws around the
periphery of the panel.
6. Loosen the thumbscrew (A) at the top edge of the panel to remove it.