Display Controls - GE Datex-Ohmeda Aisys Carestation Reference Manual

Anesthesia machine
Table of Contents

Advertisement

Aisys

1.6 Display controls

7
6
5
1.
Alarm Silence key
2.
Menu keys
3.
ComWheel
4.
Normal Screen key
5.
Quick keys
6.
Timer keys
7.
MV/TV Alarms key
Figure 1-6 • Display controls
1-8
Push to silence any active, silenceable high and medium priority alarms or
to suspend/acknowledge any non-active medium or high priority alarms.
Alarm is silenced for 120 seconds or alarm is suspended for 90 seconds.
Push to show corresponding menu.
Push to select a menu item or confirm a setting. Turn clockwise or
counterclockwise to scroll menu items or change settings.
Push to remove all menus from the screen.
Push to change corresponding gas setting or ventilator setting. Turn the
ComWheel to make a change. Push the ComWheel to activate the change.
Push to start or stop the timer. Push to reset the timer back to zero.
Push to turn off the MV and TV alarms. Push again to turn the MV and TV
alarms back on.
2
4
1
2
3
04/08 M1046983