7.10 Electronic Vaporizer (eVap) Troubleshooting

7.10.1 Vaporizer
Test Results
7-78
As a general rule, proceed with electronic vaporizer fault isolation as follows:
1. Review the system error logs for vaporizer specific entries.
• Refer to Section 7.9 for procedures related to the system error log entries.
2. Within the PC service application, perform the Vaporizer Test (Section 12.10.2).
• Follow procedures below as directed by the vaporizer test results.
3. After servicing the vaporizer, rerun the Vaporizer Test to confirm any repairs made
were successful.
The Vaporizer Test will display extended diagnostic information upon completion.
If the test passes, the message
is displayed along with the data used during the check. This indicates that the vaporizer
is ready for use.
If the check encounters a problem, it will perform an automated fault isolation
procedure. With most single fault failures the automated procedure will identify the
failed component or a small set of possibilities.
In the event of a failure, a variety of messages are possible:
• Refer to the table in Section 7.10.2 for corrective actions.
PASSED: Electronic Vaporizer Subsystem Check
04/08 M1046983