12.7.3 Power Diagnostics
M1046983 04/08
Mixer Board Power
Mixer Board Power
Label
Mixer 10VA Volts
12.5 V
5.5V
3.3V CPU
2.5V ADC Ref
Value Format
Units
OK, Fail
XX.X
Vdc
X.XX
Vdc
X.XX
Vdc
X.XX
Vdc
12 Service Application
Range
11.8 to 13.0
5.39 to 5.61
3.22 to 3.38
2.47 to 2.53
12-15