7.3 Troubleshooting Startup Screen (POST) messages - for HPDU

Message
***ERROR:
CPU data cache fault.
***ERROR:
No bootable device available.
***ERROR:
Program load failed - CRC.
***ERROR:
RAM memory error.
***ERROR:
System reset: ECxx xx xx . . .
***ERROR:
System reset: FFFF FF FF
***ERROR:
watchdog circuit failed.
***NOTE:
Alarm speaker not detected.
Check connection.
***NOTE:
CMOS battery is weak.
Please replace.
***NOTE:
RTC date/time error.
Battery may be weak.
7-4
If the Aisys system encounters a problem at startup to where it cannot initiate system
software, a BIOS error message indicating the failure will be displayed.
What it indicates
This indicates a hardware failure.
This indicates a problem with the
internal CF card.
This usually indicates a software file
corruption.
This indicates a hardware failure.
This usually indicates a software
failure.
Indicates the HPDU lithium battery
has lost contact with the holder or the
battery is below voltage.
This indicates a hardware failure.
Service is required to correct a faulty
connection to the speaker.
Service is required to replace the CPU
battery.
Service is required to replace the CPU
battery.
Troubleshooting Action Required
Replace the HPDU CPU board.
Check or replace the internal compact flash
card.
Reload the software and check out the system.
Replace the HPDU CPU board.
Report this error, along with the machine logs,
to Technical Support.
Reload software. If problem persists, replace
the internal flash card and reload software. If
problem persists, replace the HPDU CPU board.
Replace the lithium battery if below rated
voltage.
Install the software downloader card to restore
system setups.
Replace the HPDU CPU board.
Reconnect the speaker if possible. Replace the
HPDU CPU board if speaker connection can not
be corrected.
Replace the battery on the HPDU CPU board.
Reload software and check out the system.
Replace the battery on the HPDU CPU board.
Reload software and check out the system.
04/08 M1046983