12.8.3 Gas Delivery Subsystem
Mixer Pressure and Temperature
Label
O2 Pressure
O2 Pressure Cal
Balance Pressure
Balance Pres Cal
Mixer Output Pres
Mixer Output Pres Cal
O2 Temp
O2 Temp Volts
Balance Temp
Balance Temp Volts
M1046983 04/08
Mixer Pressure and Temperature
Value
Units
Range (kpa/C)
XX.XX
62.05 to 220.63
X.XXX
Vdc
0.55 to 2.90
XX.XX
62.05 to 220.63
X.XXX
Vdc
0.55 to 2.90
XX.XX
62.05 to 199.95
X.XXX
Vdc
0.55 to 2.90
XX.X
5.00 to 50.00
X.XXX
Vdc
0.25 to 3.00
XX.X
5.0 to 50.0
X.XXX
Vdc
0.25 to 3.00
12 Service Application
Range (psi/F)
Range (bar/C)
9.00 to 32.00
0.62 to 2.21
0.55 to 2.90
0.55 to 2.90
9.00 to 32.00
0.62 to 2.21
0.55 to 2.90
0.55 to 2.90
9.00 to 29.00
0.62 to 2.00
0.55 to 2.90
0.55 to 2.90
41.00 to 122.00
5.00 to 50.00
0.25 to 3.00
0.25 to 3.00
41.0 to 122.0
5.0 to 50.0
0.25 to 3.00
0.25 to 3.00