Error Messages

Error messages indicate the potential cause of a system failure. They are shown on the touch screen
panel or acquisition workstation. (Refer to Table 3-9.)
Table 3-9 Error Messages
Error
Description (in screen)
Number
1
Initialize EEPROM in Service Mode
2
EEPROM could not be initialized
3
Move without Order: Call Service
4
Fatal RAM ERROR: Call Service
5
Movement Order active during
Start-up
6
Security: Call Service
7
Initial Gauge Check: Restart
8
Communication Lost
9
Inconsistent Relay Input
10
Order without move
11
Potentiometer
12
Calibration
13
External
14
Potentiometer out of Range
15
Movements locked
16
Move Button Released
17
Incompatible Movements
18
Opposite Move
Definium 5000 X-Ray System
Operator Manual
Turn the unit Off and On. if error remains, call field
service.
Press the RESET ERROR button on the Touch Screen
Console.
Turn the Unit OFF and call field service.
Turn the Unit OFF/ON, if error remains, call field
service.
Press the Safety Lock button
Operator stopped pressing the Auto positioning
button. Press and hold until auto positioning has
finished.
A movement button has been pressed while auto
positioning was in progress. Release the movement
button.
An opposite movement button has been pressed
while other movement was in progress. Release the
opposite movement button.