Invenia ABUS 2.0 Troubleshooting and System Messages

Troubleshooting
Symptom
Blank Screen
If the system powers on with an error on the
screen.
Mechanical Noise / Thumping
Multi-Axis Lock Stops Working
System Overheating
Image doesn't transfer from the Invenia
ABUS 2.0 system
Lines in the Image
Intermittent Operation, System Lock-Ups,
System Slow-Downs, Overheating, Poor
Image Quality
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Troubleshooting and System
Table 7-2:
Invenia ABUS 2.0 Messages

Troubleshooting and System Messages

Action to Take
Troubleshoot the following:
• Is the system plugged in?
• Is the system turned on and the power button lit (green)?
• Does the transducer initialize (move across the face of the
Scan Head Assembly)?
Reboot the system, up to three (3) times. If the error still
appears, contact GE Service.
Contact GE Service.
DO NOT spray any spray or oil, or apply any grease to any
part of the system.
Contact GE Service.
Clean the filter.
CAUTION: DO NOT operate the system without an air filter.
Confirm that the Workstation Network Indicator light is
green. Shut down both the Invenia ABUS 2.0 and
Workstation; then 1st reboot the Workstation, and 2nd
reboot the Invenia ABUS 2.0.
Remove the transducer, then reverse it 180 degrees and
plug it back in.
If you still see lines on the image, but on the opposite side
of the screen, then the transducer is defective.
If the lines are still on the same side of the screen, then the
issue is on the adcquisition board or the TIP board.
In either case, contact GE Service.
Clean the system filter and let the system cool down for 30
minutes to 1 hour. Keep the system a distance of 18 inches
(~0.46 m) from curtains. CAUTION: DO NOT operate the
system without a filter.