Communication Protocol - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

COMM
EXIT

Communication Protocol

N-595
5. Press the COMM softkey.
6. Use the ADJUST UP and ADJUST DOWN buttons to select the desired baud
rate.
7. Press the EXIT softkey set the baud rate. The baud rate setting will be in
effect until the monitor is powered off.
Note:
The baud rate setup for the monitor may be saved as institutional default
settings. See Setting Institutional Defaults (Sample) on page 47. Use care
when performing this procedure because all settings will be saved as
institutional default settings.
The COMM softkey is used to select from eight communication protocols
supported by the data port. The selections are:
ASCII used for printouts
OXINET to enable communication with Oxinet II
CLINICAL intended for Nellcor use only
GRAPH for graphic printouts
AGILENT interfaces the N-595 with an Agilent (HP) monitor
SPACELBS interfaces the N-595 with a SpaceLabs monitor
MARQ interfaces the N-595 with a GE Marquette monitor
DATEX interfaces the N-595 with a Datex-Ohmeda AS/3 monitor
Note:
Selecting AGILENT, SPACELBS, MARQ, or DATEX automatically sets
the baud rate to the rate applicable for that protocol.
To change the communication protocol:
Data Port Interface Protocol
107