Simple Repairs; Exchanging A Flow Cell - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector

Simple Repairs

On the following pages repairs are described that can be carried out without opening
the main cover.
Table 23 Simple Repairs
Procedure
Flow cell exchange
Flow cell flushing
Leak sensor drying
Leak handling System
replacement

Exchanging a Flow Cell

When required
Tools required
Parts required
DO NOT install the inlet capillary to the outlet connection of the flow cell. This will result in poor
performance.
124
1100 Series FD Reference Manual
Typical Frequency
If application requires a different flow cell type or if
defective.
If flow cell is contaminated
If leak has occurred.
If broken or corroded.
If an application needs a different type of flow cell or the flow cell is
defective ( leaky ).
Two 1/4 inch wrenches for capillary connections
Standard flow cell, 8 µl, 20 bar, G1321-60005
Cuvette for off-line measurements, 8 µl, 20 bar, G1321-60007, refer to
Ho to use the Cu ette
Notes
Complete Assembly
A wavelength calibration check should
be performed after replacement.
If the flow cell is removed and
inserted, then a quick calibration
check is performed. If this fails, you
must do a wavelength recalibration,
Wa ele gth Verifi atio a d
see
Cali ratio
on page 105.
Check for leaks.
Check for leaks.
on page 140 for more information on usage.