[SENSe[1]]|SENSe2:CORRection:GAIN[1|2] Node

This node is used to provide a simple correction to the measurement. It is
possible to apply two corrections, one for the calibration factor
(CFACtor|GAIN1) and one for any other channel offset (GAIN2).
Note
Gain and loss correction are a coupled system. This means that a gain set
by [SENSe[1]]|SENSe2:CORRection:GAIN2 is represented in the
[SENSe[1]]|SENSe2:CORRection:LOSS2? command.
LOSS2 is coupled to GAIN2 by the equation:
Loss
Gain
The commands associated with this node are:
[SENSe[1]]|SENSe2:CORRection:CFACtor:GAIN[1|2][:INPut]
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe <Boolean>
HP E4418B/E4419B Programming Guide
The SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN1
command can be used for the HP 8480 series power sensors when
no sensor calibration table has been set up. If a sensor calibration
table is selected the error -221, "Settings Conflict" occurs.
1
when the default unit is linear, and
=
------------ -
Gain
when the default unit is logarithmic.
=
Loss
[:MAGNitude] <numeric_value>
[SENSe[1]]|SENSe2:CORRection:GAIN[1|2] Node
SENSe Subsystem
9-25