Editing Frequency Dependent Offset Tables - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Power Meter Remote Operation
Using Frequency Dependent Offset Tables

Editing Frequency Dependent Offset Tables

It is not possible to create any additional frequency dependent offset
tables. However, the 10 existing ones can be edited using the MEMory
subsystem. To do this:
1. Select one of the existing tables using:
MEMory:TABle:SELect <string>.
For information on naming frequency dependent offset tables see
"Naming Frequency Dependent Offset Tables", on page 1-44. For
information on the current names which you can select refer to
"Listing the Frequency Dependent Offset Table Names", on
page 1-43.
2. Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value>
{,<numeric_value>}
3. Enter the offset factors as shown in the table below using:
MEMory:TABle:GAIN <numeric_value>
{,<numeric_value>}.
Frequency
Frequency 1
Frequency 2
Frequency n
4. If required, rename the frequency dependent offset table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string>
parameter identifies the existing table name, and the second
identifies the new table name.
1-42
"
HP E4418B/E4419B Programming Guide
Offset
Offset 1
Offset 2
"
Offset n