Parameters
Item
numeric_value
1. The following measurement units can be used:
2. All frequencies are truncated to a multiple of 1 kHz.
Example
Query
MEMory:TABLe:FREQuency?
The query returns a list of frequency points for the table currently
selected. The frequencies are returned in Hz.
Query Example
HP E4418B/E4419B Programming Guide
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
Description/Default
A numeric value for the frequency. The
default units are Hz.
MEM:TABL:FREQ 200kHz,600kHz
MEM:TABL:FREQ?
Hz
3
kHz (10
)
6
MHz (10
)
9
GHz (10
)
This command enters
frequencies of 200 kHz and
600 kHz into the currently
selected table.
This command queries the
frequency points in the currently
selected table.
MEMory Subsystem
Range of Values
1 kHz to
1, 2
999.9 GHz