MEASure[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right.
The parameter DEFault is used as a place holder. Specifying
parameter value unchanged.
2. When the measurement result is linear this parameter represents the number
of significant digits. When the measurement result is logarithmic 1 to 4
represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
MEAS2:POW:AC? -70DBM,1,(@1) This command queries the lower
HP E4418B/E4419B Programming Guide
Item
Description/Default
source list
This is the channel on which the
command is implemented.
If unspecified the current window
setup is used. However, on the
HP E4419B, if the window shows a
ratio or difference measurement,
the upper window defaults to
channel A and the lower window to
channel B.
DEF
means DEFault. This is not equivalent to the DEFault
MEASurement Instructions
Range of Values
(@1)
(@2) (HP E4419B
only)
DEF
window's measurement of
channel A, using an expected power
level of -70 dBm and a resolution
setting of 1.
leaves the
2-51