The Channel A and B sensor connected bits (bits 1 and 2), when queried
with the STATus:DEVice:CONDition? query are set to:
1, when a power sensor is connected.
0, when no power sensor is connected.
The Channel A and B sensor connected bits (bits 1 and 2), when queried
with the STATus:DEVice:EVENt? query indicate whether a power sensor
has been connected or disconnected depending on the state of the
corresponding bits of STATus:DEVice:NTRansition and
STATus:DEVice:PTRansition. If the corresponding bit in:
STATus:DEVice:NTRansition is 1, then
STATus:DEVice:EVENt? is set when a power sensor is
disconnected. Note, querying STATus:DEVice:EVENt? clears the
STATus:DEVice:EVENt?register.
STATus:DEVice:PTRansition is 1, then
STATus:DEVice:EVENt? is set when a power sensor is connected.
Note, querying STATus:DEVice:EVENt? clears the
STATus:DEVice:EVENt?register.
The Channel A and B sensor error bits (3 and 4) are set to:
1, if the HP E-series power sensor EEPROM has failed or if there
are power sensors connected to both the rear and front panel
connectors.
0, for every other condition.
The Channel A and B sensor Front/Rear bits (bits 5 and 6) are set to:
1, if the power sensor is connected to the rear panel.
0, if the power sensor is connected to the front panel.
The Front Panel key press bit (bit 14), when queried with the
STATus:DEVice:EVENt? query indicates whether any front panel keys
have been pressed since power up or since you last queried the device
status register. Note, this bit ignores the :NTRansition, and
:PTRansition registers,and a:CONDition?query always returns a
0.
HP E4418B/E4419B Programming Guide
STATus Subsystem
Device Status Register Sets
10-9