SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
Note
Syntax
SENS
SENS2
Example
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether
automatic filter length is enabled or disabled.
Query Example
9-8
For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best
mode of operation. However, manual filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>) is
useful in applications requiring either high resolution or fast
settling times.
:
AVER
1
AVER:COUN:AUTO OFF
1 is returned when automatic filter length is enabled.
0 is returned when automatic filter length is disabled.
AVER:COUN:AUTO?
:COUN
:AUTO
This command disables
automatic filter length selection for
channel A.
This command queries whether
automatic filter length selection is
on or off for channel A.
HP E4418B/E4419B Programming Guide
Space
0|OFF
1|ON
ONCE
?

Hide quick links: