Parameters
Item
numeric_value
Example
Reset Condition
On reset, the speed is set to 20 readings/second.
Query
[SENSe[1]]|SENSe2:SPEed?
The query returns the current speed setting, either 20, 40 or 200.
Query Example
Error Message
If <speed> is not 20, 40 or 200, error -224, "Illegal parameter value"
occurs.
If an HP E-series power sensor is not connected and the
<numeric_value> is set to 200 readings/second the error -241,
"Hardware missing" occurs.
HP E4418B/E4419B Programming Guide
Description/Default
A numeric value for the measurement
speed. The default is 20.
SPE 40
SPE?
[SENSe[1]]|SENSe2:SPEed <numeric_value>
This command sets the channel A
speed to 40 readings/second.
This command queries the current
speed setting for channel A.
SENSe Subsystem
Range of Values
20
40
200
9-55