Status:operation:ulfail[:Summary] - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

STATus:OPERation:ULFail[:SUMMary]

The operation status upper limit fail summary register set contains
information on the upper limit fail status of the power meter.
The following bits in these registers are used by the power meter:
The appropriate bits are set if a channel upper limit test fails or a window
upper limit test fails.
These bits are cleared if a measurement is made and the test is enabled
and passes.
Syntax
HP E4418B/E4419B Programming Guide
Bit
Decimal
Definition
Number
Weight
0
-
Not used
1
2
Channel A ULFail Status
2
4
Channel B ULFail Status (HP E4419B only)
3
8
Upper window ULFail Status
4
16
Lower window ULFail Status
5-15
-
Not used
:OPER
STAT
STATus:OPERation:ULFail[:SUMMary]
:ULF
:SUMM
STATus Subsystem