MEASurement Instructions
CONFigure[1|2][:SCALar][:POWer:AC] [<expected_value>[,<resolution>[,<source list>]]]
Parameters
Refer to "The Optional Parameters", on page 2-3 for additional details on
the parameters in this command.
Item
expected_value
(for the
expected
power level)
resolution
source list
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in the
specified order. If parameters are omitted, they will default from the right. The
parameter DEFault is used as a place holder. Specifying
value unchanged.
2. When the measurement result is linear this parameter represents the number of
significant digits. When the measurement result is logarithmic 1 to 4 represents a
resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
2-10
Description/Default
A numeric value for the expected
power level. The units of
measurement are dBm and W. The
default units are defined by
UNIT:POWer.
A numeric value for the resolution.
If unspecified the current
resolution setting is used.
This is the channel on which the
command is implemented.
If unspecified the current window
setup is used. However, on the
HP E4419B, if the window shows a
ratio or difference measurement,
the upper window defaults to
channel A and the lower window to
channel B.
means DEFault. This is not equivalent to the DEFault
DEF
CONF1:POW:AC DEF,2,(@1)
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
(@1)
(@2) (HP E4419B only)
leaves the parameter
DEF
This command configures the
upper window to measure the
power of channel A, using the
current sensor range and a
resolution setting of 2.
HP E4418B/E4419B Programming Guide