SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
Error Messages
If <string> is not valid, error -224, "Illegal parameter value" occurs.
If a table called <string> does not exist, error -256, "File name not
found" occurs.
When a sensor calibration table is selected, the power meter verifies that
the number of calibration points defined is one more than the number of
frequency points defined. When a frequency dependent offset table is
selected, the power meter verifies that the number of offset points defined
is equal to the number of frequency points defined. If this is not the case,
error -226, "Lists not the same length" occurs.
If the CSET1 command is used when an HP E-series power sensor is
connected the error -241, "Hardware missing" occurs.
9-16
CORR:CSET1?
This command queries the sensor
calibration table currently used for
channel A.
HP E4418B/E4419B Programming Guide