IDLE
STATE
TRIG:SOUR IMM
TRIG:SOUR BUS
TRIG:SOUR HOLD
SEQUENCE
OPERATION
STATE
Idle State
Turning power on, sending an HP-IB CLEAR, sending a *RST or an
:ABORt forces the trigger system into the idle state. The trigger system
remains in the IDLE state until it is initiated by INITiate:CONTinuous
ON or INITiate:IMMediate. Once one of these conditions is satisfied the
trigger system moves to the initiate state.
HP E4418B/E4419B Programming Guide
Figure 1-8: Trigger System
:ABORt
*RST
INIT[:IMM]
INIT:CONT ON
YES
Wait - for - trigger
TRIG:SOURce
Wait
TRIGGERED
Power Meter Measurement Actions
Idle State
NO
or
YES
TRIG:IMMediate
state
TRIG:DEL
Power Meter Remote Operation
Triggering the Power Meter
NO
INITIATE
STATE
Is
INIT:CONT ON
EVENT
DETECTION
STATE
1-63