Power Meter Remote Operation
Making Measurements
The following program segment assumes that channel A is currently being
measured on the upper window.
ABOR1
CONF1 DEF,3
INIT1
FETC1?
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off.
ABOR1
CONF1 DEF,3
TRIG1:DEL:AUTO OFF
INIT1:IMM
FETC1?
1-26
Aborts channel A.
Configures the upper window to
make a measurement using the
current setting of the expected
power and source list and a
resolution setting of 3.
Causes channel A to make a
measurement.
Retrieves the upper window's
measurement.
HP E4418B/E4419B Programming Guide