Calculate[1|2]:Relative:state <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

CALCulate Subsystem
CALCulate[1|2]:RELative:STATe <Boolean>
CALCulate[1|2]:RELative:STATe <Boolean>
This command is used to enable and disable relative mode. If the
command is:
Syntax
Example
Reset Condition
On reset, relative mode is disabled.
Query
CALCulate[1|2]:RELative:STATe?
The query returns a 1 or 0 into the output buffer.
Query Example
Error Message
If CALCulate:RELative:STATe is set to ON while SENSe:SPEed is set to
200, the error -221, "Settings Conflict" occurs.
3-28
disabled, the measurement signal remains unchanged.
enabled, the current relative value set by
CALCulate:RELative:MAGnitude:AUTO is applied to the
measurement signal.
CALC
1
:REL
2
CALC1:REL:STAT OFF
1 is returned when relative mode is enabled.
0 is returned when relative mode is disabled.
CALC1:REL:STAT?
Space
:STAT
This command disables the
relative mode on the upper
window.
This command queries whether
relative mode is off or on.
HP E4418B/E4419B Programming Guide
0|OFF
1|ON
?