HP E4418B Programming Manual page 194

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right.
The parameter DEFault is used as a place holder. Specifying
parameter value unchanged.
2. When the measurement result is linear this parameter represents the number
of significant digits. When the measurement result is logarithmic 1 to 4
represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
MEAS1:DIFF:REL? DEF,3,(@2),(@1)
HP E4418B/E4419B Programming Guide
MEASure[1|2][:SCALar][:POWer:AC]:DIFFerence:RELative?
Item
Description/Default
source list
This channel list specifies between
which channels the difference is
calculated.
If unspecified and the current
window setup is a difference
measurement then this difference
setup is used, otherwise it defaults
to channel A-B.
means DEFault. This is not equivalent to the DEFault
DEF
[<expected_value>[,<resolution>[,<source list>]]]
This command queries the
upper window's relative
difference measurement of
channel B - channel A,
using the current sensor
range and a resolution
setting of 3 on both
channels.
MEASurement Instructions
Range of Values
(@1),(@2)
(@2),(@1)
leaves the
DEF
2-57