Display[:Window[1|2]]:Resolution <Numeric_Value - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

DISPlay Subsystem
DISPlay[:WINDow[1|2]]:RESolution <numeric_value>
DISPlay[:WINDow[1|2]]:RESolution <numeric_value>
This command is used to set the resolution of the measurement result in
the selected window.
When the measurement result is linear this parameter represents the
number of significant digits. When the measurement result is logarithmic
1 through 4 represents a resolution of 1, 0.1, 0.01, and 0.001 respectively.
The resolution set for the measurement window also affects the level of
averaging in the SENSe subsystem when auto averaging is selected. Also,
as a measurement from a selected channel may be displayed in more that
one window (either directly as a single channel measurement or with a
ratio or difference measurement (HP E4419B only)) there is a defined
relationship between the window resolution and the resolution used to
determine the auto averaging level. The resolution level used to determine
auto averaging for a selected channel is the greater of the resolutions set
for all windows in which the channel is displayed.
Syntax
DISP
Parameters
Item
numeric_value
5-14
:WIND
1
:RES
2
Description/Default
A numeric value for the window
resolution. DEF is 3, MIN is 1, and MAX
is 4.
Space
numeric_value
DEF
MIN
MAX
Space
?
Range of Values
1
2
3
4
DEF
MIN
MAX
HP E4418B/E4419B Programming Guide
MIN
MAX

Hide quick links: