Reviewing Table Data
To review the data stored in a sensor calibration table, use the following
commands:
MEMory:TABLe:SELect "Sense1"
Select the sensor calibration table named "Sense1".
MEMory:TABLe:SELect?
Query command which returns the name of the currently selected
table.
MEMory:TABLe:FREQuency:POINTs?
Query command which returns the number of stored frequency
points.
MEMory:TABLe:FREQuency?
Query command which returns the frequencies stored in the sensor
calibration table (in Hz).
MEMory:TABLe:GAIN[:MAGNitude]:POINTs?
Query command which returns the number of calibration factor
points stored in the sensor calibration table.
MEMory:TABLe:GAIN[:MAGNitude]?
Query command which returns the calibration factors stored in the
sensor calibration table. The first point returned is the reference
calibration factor.
Modifying Data
If you need to modify the frequency and calibration factor data stored in a
sensor calibration table you need to resend the complete data lists. There
are two ways to do this:
1. If you have retained the original data in a program, edit the
program and resend the data.
2. Use the query commands shown in "Reviewing Table Data", on
page 1-37 to enter the data into your computer. Edit this data,
then resend it.
HP E4418B/E4419B Programming Guide
Power Meter Remote Operation
Using Sensor Calibration Tables
1-37