Output:roscillator[:State] <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

OUTPut:ROSCillator[:STATe] <Boolean>
This command is used to switch on and off the POWER REF output
Syntax
Example
Reset Condition
On reset, the POWER REF output is switched off.
Query
OUTPut:ROSCillator[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the POWER REF is switched on.
Query Example
HP E4418B/E4419B Programming Guide
OUTP
:ROSC
OUTP:ROSC:STAT 1
1 is returned when the POWER REF output is switched ON.
0 is returned when the POWER REF output is switched OFF.
OUTP:ROSC?
OUTPut:ROSCillator[:STATe] <Boolean>
Space
:STAT
This command switches on the
POWER REF output.
This command queries the status
off the POWER REF output.
OUTput Subsystem
0|OFF
1|ON
?
8-3