[Sense[1]]|Sense2:Correction:cset[1]|Cset2:State <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <Boolean>
This command is used to enable and disable the use of the currently active
sensor calibration (CSET1) or frequency dependent offset (CSET2) table.
When a table has been selected and enabled, the calibration factors/offsets
stored in it can be used by specifying the required frequency using the
[SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is
taken from the sensor calibration table and is used during calibration.
Syntax
SENS
1
SENS2
Example
Reset Condition
On reset, the sensor calibration table and frequency dependent offset
table are not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a
table is enabled or disabled.
HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <Boolean>
:
CORR
:CSET
:CSET2
CORR:CSET1:STAT 1
1 is returned when the table is enabled.
0 is returned when the table is disabled.
Space
1
:STAT
This command enables the use of
the currently active sensor
calibration table for channel A.
SENSe Subsystem
0|OFF
1|ON
?
9-17