HP E4418B Programming Manual page 307

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>
Parameters
Item
numeric_value
Example
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values
associated with MIN and MAX. The format of the response is <NR1>.
Query Example
Error Message
If a filter length value is entered using
[SENSe[1]]|SENSe2:AVERage:COUNt while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings
Conflict" occurs. However, the filter length value is set but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically
set ON.
9-6
Description/Default
A numeric value defining the filter length.
The default value, DEF, is 4. The value for
MIN is 1. The value for MAX is 1024.
AVER:COUN 400
AVER:COUN?
Range of Values
1 to 1024
DEF
MIN
MAX
This command enters a filter
length of 400 for channel A.
This command queries the filter
length for channel A.