STATus Subsystem

STATus:OPERation:CALibrating[:SUMMary]

STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains
information on the calibrating status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO
ONCE) and at the beginning of calibration (CALibration:AUTO ONCE).
Also for the compound command/query CALibration[:ALL]?, this bit is
set at the beginning of the calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
10-12
Bit
Decimal
Definition
Number
Weight
0
-
Not used
1
2
Channel A CALibrating Status
2
4
Channel B CALibrating Status (HP E4419B only)
3-15
-
Not used
STAT
:OPER
:CAL
:SUMM