Calibration[1|2]:Rcalibration <Boolean - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

CALibration Subsystem
CALibration[1|2]:RCALibration <Boolean>
CALibration[1|2]:RCALibration <Boolean>
This command enables and disables the zero/cal lockout facility. With the
lockout facility enabled the power meter is prevented from making
measurements until the sensor connected has been zeroed and calibrated.
Syntax
Example
Reset Condition
On reset, the state of the zero/cal lockout is unaffected.
Query
CALibration[1|2]:RCALibration?
The query enters a 1 or 0 into the output buffer indicating whether
zero/cal lockout is enabled or disabled.
Query Example
4-10
CAL
1
2
CAL1:RCAL 1
1 is returned if zero/cal lockout is enabled.
0 is returned if zero/cal lockout is disabled.
CAL1:RCAL?
Space
:RCAL
?
This command enables the zero
/cal lockout facility on channel A.
This command queries whether or
not the zero/cal lockout facility is
enabled for channel A.
HP E4418B/E4419B Programming Guide
0|OFF
1|ON

Hide quick links:

Advertisement

Table of Contents
loading