SENSe Subsystem
[SENSe[1]]|SENSe2:LIMit:STATe <Boolean>
[SENSe[1]]|SENSe2:LIMit:STATe <Boolean>
This command is used to enable and disable the test limits for the
specified channel.
Syntax
Example
Reset Condition
On reset, limit checking is disabled.
Query
[SENSe[1]]|SENSe2:LIMit:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
limits testing feature for the specified channel.
Query Example
Error Message
If [SENSe[1]]|SENSe2:LIMit:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings
Conflict" occurs.
9-48
:
SENS
1
SENS2
SENS2:LIM:STAT 1
1 is returned when limits testing is enabled.
0 is returned when limits testing is disabled.
SENS1:LIM:STAT?
LIM
:STAT
This command enables the limit
checking function for channel B.
This command queries whether
the limit checking function for the
channel A is on or off.
HP E4418B/E4419B Programming Guide
Space
0|OFF
1|ON
?