HP E4418B Programming Manual page 323

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]
Query Example
Error Messages
If a duty cycle value is entered using
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error -221, "Settings
Conflict" occurs. However, the duty cycle value is set but the
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe command is
not automatically set ON.
If this command is used when an HP E-series power sensor is connected,
the error -310, "System error;Dty Cyc may impair accuracy with ECP
sensor" occurs. If you are using the HP E4419B the error message
specifies the channel.
9-22
CORR:GAIN3?
This command queries the current
setting of the duty cycle for
channel A.
HP E4418B/E4419B Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4419b

Table of Contents