Abort[1|2] - HP E4418B Programming Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

ABORt[1|2]

The ABORt command removes the specified channel from the
"wait-for-trigger" state and places it in the "idle" state. It does not affect
any other settings of the trigger system. When the INITiate command is
sent, the trigger system responds as it did before ABORt was executed.
If INITiate:CONTinuous is ON, then after ABORt the specified channel
immediately goes into the "wait-for-trigger" state.
Syntax
Example
HP E4418B/E4419B Programming Guide
ABOR
1
2
ABOR
TRIGger Subsystem
This command places channel A in
the "idle" state.
ABORt[1|2]
12-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4419b

Table of Contents