Editing Sensor Calibration Tables
It is not possible to create any additional sensor calibration tables.
However, the 20 existing ones can be edited using the MEMory subsystem.
To do this:
1. Select one of the existing tables using:
MEMory:TABle:SELect <string>.
For information on naming sensor calibration tables see "Naming
Sensor Calibration Tables", on page 1-36. For information on the
current names which you can select refer to "Listing the Sensor
Calibration Table Names", on page 1-34.
2. Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value>
{,<numeric_value>}
3. Enter the calibration factors using:
MEMory:TABle:GAIN <numeric_value>
{,<numeric_value>}. The first parameter you enter should be
the reference calibration factor, each subsequent parameter is a
calibration factor in the sensor calibration table. This means that
entries in the frequency list correspond as shown with entries in
the calibration factor list.
Frequency
Frequency 1
Frequency 2
Frequency n
4. If required, rename the sensor calibration table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string>
parameter identifies the existing table name, and the second
identifies the new table name.
HP E4418B/E4419B Programming Guide
Calibration Factor
Reference Calibration
Calibration Factor 1
Calibration Factor 2
"
Calibration Factor n
Power Meter Remote Operation
Using Sensor Calibration Tables
Factor
"
1-33