CALibration[1|2][:ALL]

Note
This command causes the power meter to perform a calibration sequence
on the specified channel. The command assumes that the power sensor is
connected to the POWER REF output. The calibration sequence consists
of:
For the HP 8480 series power sensors the reference calibration factor used
during this calibration can be derived from either an active sensor
calibration table or the value entered using CALibration:RCFactor.
The actual value used is the one which was most recently set. That is, a
value entered using CALibration:RCFactor is overridden if a sensor
calibration table is subsequently selected and enabled. Conversely,
CALibration:RCFactor overrides any reference calibration factor
previously set from a sensor calibration table. To determine the currently
set reference calibration factor use CALibration:RCFactor?
The HP E-series power sensors have their sensor calibration tables stored
in EEPROM which means that the reference calibration factor is
automatically downloaded by the power meter.
Syntax
Example
HP E4418B/E4419B Programming Guide
This command is identical to CALibration[1|2][:ALL]?,
however, unlike the query it does not provide a response to
indicate whether the calibration has been successful or not.
Zeroing the power meter (CALibration:ZERO:AUTO ONCE), and
calibrating the power meter (CALibration:AUTO ONCE).
CAL
1
:ALL
2
CAL1:ALL
CALibration Subsystem
CALibration[1|2][:ALL]
This command causes the power
meter to perform a calibration
sequence on channel A.
4-3