MEMory:CATalog:TABLe?

This command is used to list the stored sensor calibration and frequency
dependent offset tables.
The power meter returns the data in the form of two numeric parameters
and as many strings as there are stored tables.
<numeric_value>,<numeric_value>{,<string>}
For example, a sample of a response may look like:
1178,10040,"DEFAULT,TABL,14","HP8481A,TABL,116",
"HP8482A,TABL,74","HP8483A,TABL,62"...........
HP E4418B/E4419B Programming Guide
The first numeric parameter indicates the amount of memory, in
bytes, used for the storage of tables.
The second parameter indicates the memory, in bytes, available
for the storage of tables.
Each string parameter returned indicates the name, type and size
of a stored table:
<string>,<type>,<size>
<string> indicates the name of the table.
<type> indicates TABL for a table.
<size> indicates the size of the table in bytes.
MEMory Subsystem
MEMory:CATalog:TABLe?
7-7